REMI Walk in chambers are designed for simulating, monitoring & control of environmental conditions like temperature & humidity. All these chambers are double walled & modular in construction, easy to assemble at site. They are available in standard sizes & can also be tailor made to suit specific requirements. Forced air circulation with the help of fan blower maintains uniform conditions. Use of intelligent control systems & dedicated REMI Datasoft software complying to 21 CFR Part 11 requirements greatly enhances the performance. These chambers can be provided with standby refrigeration & humidity systems (optional) for uninterrupted functioning.
Technical Data
WALK-IN STABILITY CHAMBERS | WSC-50 | WSC-80 | WSC-100 | WSC-120 | WSC-150 |
WALK-IN COLD ROOM | WCR-50 | WCR-80 | WCR-100 | WCR-120 | WCR-150 |
Internal Volume (Liters) | 5000 | 8000 | 10000 | 12500 | 15000 |
Number of Shelves | 16 | 24 | 32 | 48 | 48 |
Internal Dimension W x D x H (mm) | 1250 x 2000 x 2000 | 2000 x 2000 x 2000 | 2000 x 2000 x 2500 | 2000 x 2500 x 2400 | 2500 x 2500 x 2400 |
External Dimension W x D x H (mm) | 1410 x 2950 x 2160 | 2160 x 2950 x 2160 | 2160 x 2950 x 2660 | 2160 x 3450 x 2560 | 2660 x 3450 x 2560 |
Internal Body Material | Stainless Steel 304 grade (Standard Models), Stainless Steel 316 grade (GMP Models) | ||||
External Body Material | Powder Coated CRCA Steel (Standard Models), Stainless Steel 304 grade (GMP Models) | ||||
Insulation | 80 mm minimum, CFC free polyurethane foam | ||||
Illumination | Fluorescent Tubes | ||||
Control & Display | Microprocessor based PID control system with large 4" LCD Display | ||||
Power Failure Alarm | Audio Visual Alarm | ||||
Door Open Alarm | Audio Alarm in case door open for over one minute | ||||
Electrical | 220-240 volts, 50Hz, Single Phase |